Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.14365/6253
Full metadata record
DC FieldValueLanguage
dc.contributor.authorSavas, Sumeyra-
dc.contributor.authorKilic, Yalin-
dc.contributor.authorGharibzahedi, Seyed Mohammad Taghi-
dc.contributor.authorAltintas, Zeynep-
dc.date.accessioned2025-06-25T18:05:58Z-
dc.date.available2025-06-25T18:05:58Z-
dc.date.issued2025-
dc.identifier.issn2214-1804-
dc.identifier.urihttps://doi.org/10.1016/j.sbsr.2025.100822-
dc.description.abstractA graphene quantum dot (GQD)-based nanozyme-enhanced electrochemical sensor was developed for the ultra-sensitive and rapid Staphylococcus aureus detection in PBS, undiluted milk, and blood serum. The S. aureus concentration for immunosensor calibration was electrochemically determined to be 1.24x 10(9) CFU mL(-1) (R-2 = 0.99), utilizing an innovative smartphone-based electrochemical device. Transmission electron (TEM) and atomic force (AFM) microscopy analyses of GQDs revealed uniform nanoparticles (8-10 nm) with surface heights of 2 nm and 21 nm at different scan areas. Fluorescence and Fourier transform infrared spectroscopy spectra showed an emission at 530 nm with an excitation wavelength of 485 nm, along with the presence of hydroxyl, carboxyl, and aromatic groups on the GQD surfaces. The maximum current peak for GQDs was obtained by optimizing electrochemical properties through amperometry at +0.9 V. The optimal signal intensity- and limit of detection (LOD)-based GQD concentrations were 0.0125, 0.5, and 1 mg mL(-1) for PBS, undiluted milk, and blood serum, respectively. The optimal antibody concentration for S. aureus detection was 25 mu g mL(-1), resulting in a 91 % amperometric suppression and a 94 % fluorescence quenching. The developed biosensor effectively detected S. aureus, with calculated LODs of 1, 4, and 344 CFU mL(-1) in PBS, undiluted milk, and blood serum, respectively. TEM, AFM, and contact angle analyses confirmed S. aureus binding to the GQD-antibody bioconjugate, shifting surface height to 35 nm and contact angle from similar to 30 degrees to similar to 65 degrees. The immunosensor showed high specificity for S. aureus with minimal cross-reactivity to Listeria monocytogenes, Staphylococcus epidermidis, Enterococcus faecium, and Streptococcus pneumoniae.en_US
dc.description.sponsorshipBandirma Onyedi Eylul University Scientific Research Foundation (BAP) [24-1004-004]; Scientific and Technical Research Council of Turkey (TUBITAK) [1059B192301157]; German Research Foundation [428780268, 541301469, INST 257/739, INST 257/742-1]; Aventis Foundation [80304368]; Joachim Herz Foundation [803043]; European Union [101097989]; Horizon Europe - Pillar III [101097989] Funding Source: Horizon Europe - Pillar IIIen_US
dc.description.sponsorshipThe authors gratefully acknowledge the support provided by the Bandirma Onyedi Eyluel University Scientific Research Foundation (BAP, Grant number: 24-1004-004) and the Scientific and Technical Research Council of Turkey (TUBITAK) for providing a scholarship (Grant number: 1059B192301157). The authors also acknowledge the German Research Foundation (Project numbers: 428780268, 541301469, INST 257/739, and INST 257/742-1), the Aventis Foundation (Project number: 80304368), Joachim Herz Foundation (Project number: 803043) and the European Union (Project number: 101097989) for their financial support.en_US
dc.language.isoenen_US
dc.publisherElsevier B.V.en_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectAmperometric Detectionen_US
dc.subjectFoodborne Pathogensen_US
dc.subjectGraphene Quantum Dotsen_US
dc.subjectSmartphone-Based Electrochemical Immunosensoren_US
dc.subjectStaphylococcus Aureusen_US
dc.titleA Novel Smartphone-Based Nanozyme-Enhanced Electrochemical Immunosensor for Ultrasensitive Direct Detection of Staphylococcus Aureus in Milk and Blood Serumen_US
dc.typeArticleen_US
dc.identifier.doi10.1016/j.sbsr.2025.100822-
dc.identifier.scopus2-s2.0-105007459339-
dc.departmentİzmir Ekonomi Üniversitesien_US
dc.authorwosidAltintas, Zeynep/O-3429-2015-
dc.authorwosidGharibzahedi, Seyed Mohammad Taghi/S-1190-2017-
dc.authorscopusid57194141784-
dc.authorscopusid57212488581-
dc.authorscopusid36486238300-
dc.authorscopusid54404684100-
dc.identifier.volume49en_US
dc.identifier.wosWOS:001513281100001-
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.scopusqualityQ1-
dc.identifier.wosqualityN/A-
dc.description.woscitationindexEmerging Sources Citation Index-
item.cerifentitytypePublications-
item.grantfulltextnone-
item.languageiso639-1en-
item.fulltextNo Fulltext-
item.openairetypeArticle-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
crisitem.author.dept05.02. Biomedical Engineering-
Appears in Collections:Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection
Show simple item record



CORE Recommender

SCOPUSTM   
Citations

4
checked on Oct 29, 2025

WEB OF SCIENCETM
Citations

4
checked on Oct 29, 2025

Page view(s)

80
checked on Nov 3, 2025

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.